Sumber Aneka Karya Abadi - Your trusted partner for laboratory instrument

Search

THERMO ORION SCIENTIFIC, Orion START9402 all-in-one titrator standard ROSS kit

Prod #START9402

Gunakan titrator all-in-one Orion Star T940 untuk titrasi konsentrasi ion, redoks, dan pH yang fleksibel termasuk titrasi titik ekivalen, titrasi titik akhir pH atau mV yang telah ditetapkan ditambah mode beberapa penambahan yang diketahui (MKA) untuk penambahan berbagai ion yang diketahui secara otomatis.

Product Overview

Calibration Points (ISE) 1 to 5
Calibration Points (pH) 1 to 5
Data Points 100 data sets each: sample titration, titrant standardization, calibration and direct measure
For Use With (Application) Acid-base titrations, redox titrations, ion concentration titrations
IP Rating IP-51
Languages Chinese, English, French, German, Italian, Japanese, Korean, Portuguese and Spanish
Material ABS Plastic
Model Orion Star T940
Range (ISE) 0.0001 to 19990
Range (mV) -2000.0 to +2000.0 mV
Range (pH) -2 to 20
Resolution (ISE) 0.0001 minimum, 1 to 4 significant digits (user selectable)
Resolution (pH) 0.001, 0.01, 0.1 (user selectable)
Temperature Range (Metric) -5.0°C to +100.0°C
Burette Functions Automatic rinse cycles for flushing and discrete dispense volume with continuous option; Exceeds acc
Display Type 5.7” color capacitive touchscreen, 640 x 480 resolution
Functions Available Sample titration, titrant standardization, calibration and direct measure
Type Acid/base, redox and ion titrator
Description Orion Star T940 all-in-one titrator ROSS Sure-Flow kit
Includes 8172BNWP ROSS Sure-Flow pH electrode, 927007MD ATC probe, 20 mL burette, stirrer probe, dispenser probe, standard tubing kit, 1 L plastic bottle, GL38 bottle cap with two ports and drying tube, computer cable, USB drive, 110-240 V power supplies
Advanced ISE Functions Multiple known addition (MKA); automated calibration and calculation of sample concentration by adding aliquots of standard to sample using the burette
Titration Technique Equivalence point or preset endpoint
Equivalence Points 1 to 2
Preset Endpoints 1 to 3
Titration Types Direct or back titration
Blank Options Fixed value or variable value using titration
Cycles Per Titration Run 1 to 5 cycles, option to exclude cycle results from average results and RSD calculations
Titrant Determination Standardization titration or manual concentration input
Titration Process Control Routine, quick, careful or user defined
Precision 0.5% RSD, dependent on environment and handling conditions
Titration Setup Wizard Yes, using Start a New Titration workflow
Burette Sizes 10 mL, 20 mL (included) or 50 mL
Resolution Burette drive offers advanced micro-step technology that provides 25,600 micro-steps per motor revolution for smooth and precise burette positioning (over 2 million micro-steps over the full burette stroke range)
Stirrer Probe Speeds 5 user selectable speeds (250 to 3700 RPM)
Methods Up to 10 methods with optional password protection; import/export methods via USB flash drive
Data Management Details Data log export as CSV or PDF file, short or long format
Sample ID Auto-incremental, manual or off
Time and Date Yes, with nonvolatile battery backup
Memory Non-volatile memory preserves data logs and settings
Updateable Firmware Yes, using USB flash drive
Alarm Output Notifications for Titration Cycle Complete, Maximum Titrant Volume, Data Log Full, Calibration Due, Maintenance Reminder
Accuracy (pH) ±0.002pH
Accuracy (ISE) ±0.2 mV or ±0.05% of reading, whichever is greater
Resolution (mV) 0.1 mV
Accuracy (mV) ±0.2 mV
Resolution (Temperature) 0.1 °C
Accuracy (Temperature) ±0.2 °C
Calibration pH, mV, ion concentration (ISE) and temperature
Calibration Points 1 point relative mV (RmV) mode
Temperature Probe Calibration Optional 1 point offset calibration
Temperature Selection Manual or automatic
Backlight Option Yes, adjustable
Channels 1
Probe Type pH electrode, redox electrode or ion selective electrode (ISE); reference half-cell electrode; ATC temperature probe; stirrer probe
Inputs BNC (pH, redox or ion selective electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe), stirrer input (stirrer probe)
Outputs 3 USB ports for printer, USB flash drive and computer software connections
Certifications/Compliance CE, TUV 3-in-1, FCC, EN/EIC61326-1, IEC 61010
Electrical Requirements 100-240V, 50/60 Hz
Dimensions (L x W x H) 10” x 16” x 14”; 25.4 cm x 40.6 cm x 35.6 cm
Temperature (English) Operating 41°F to 104°F
Temperature (Metric) Operating 5°C to 40°C
Relative Humidity Range 5 to 85%, non-condensing
Warranty 1 year replacement warranty
Unit Size Each

Resources