Spesifikasi :
|
Accuracy |
LR model: ±4 µg/L or ±4% of reading, whichever is greater; |
|
Alarm |
Four relays output; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum |
|
Certifications |
CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000) |
|
Controller Compatibility |
SC200, SC1000 |
|
Detection limit
|
LR model: 4 µg/L |
|
Enclosure Rating |
IP56 / NEMA 4X |
|
Fitting type |
Sample line and sample bypass drain: 6 mm (¼-in.) |
|
Flow rate |
55 - 300 mL/min |
|
Grab sample |
Grab Sample In and Grab Sample Out capability |
|
Light source |
Class 1M LED (light emitting diode) with a peak wavelength of 810 nm |
|
Manual Languages |
English |
|
Measuring Principle |
Colorimetric |
|
Measuring range
|
LR model: 4 - 5000 µg/L as PO4; |
|
Mounting |
Wall, panel or table |
|
Operating Humidity |
5 - 95 % non-condensing (indoor use only) |
|
Operating Temperature Range |
5 - 45 °C |
|
Output |
4 - 20 mA |
|
Number of channels |
1, 2, 4; programmable sequence |
|
Pollution Degree |
2/II |
|
Power Requirements (Hz) |
50/60 Hz |
|
Power Requirements (Voltage) |
100 - 240 V AC, 24 V DC |
|
Protection Class |
I |
|
Reagent Consumption |
2 L of each reagent every 90 days with 15 minute cycle time |
|
Relays |
Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum |
|
Repeatability |
LR model: ±1% |
|
Response Time |
Typically, 9 minutes at 25°C (77°F); changes with temperature |
|
Response Time |
Typically, 9 minutes at 25°C (77°F); changes with temperature |
|
Sample Pressure |
0.14 - 6 bar (to Preset Pressure Regulator) |
|
Sample Temperature |
5 - 50 °C |
|
Storage Conditions |
-20 - 60 °C |
|
Warranty |
2 years |
|
Weight |
20 kg without reagents and standards, 36.3 kg with reagents |