Triple Overtemperature Protection
in case of overtemperature due to failure, the
heating is switched off at approx. 10 °C above
the set temperature (fixed value)
•
independently working, digitally adjustable
electronic overtemperature controller TWW
protection class 3.1
•
resolution of display and setting accuracy: 0,5
°C below 99,9 °C, 1 °C above 100 °C
•
mechanical temperature limiter TB protection
class 1 switching the heating off at approx. 10
°C above max. oven temperature
Textured Stainless Steel Casing
•
w x h x d: 550 x 680 x 480 mm
•
fully insulated stainless steel door with double
locking and 4-point adjustment
•
rear zinc-plated steel
Interior - Heating Concept
•
w x h x d: 400 x 400 x 330 mm, 53 l
•
easy-to-clean interior, made of stainless steel,
reinforced by deep drawn ribbing with
integrated and protected large-area heating
on four sides
•
2 stainless steel grids
Temperature Range
•
from +30 °C (however, at least 10 °C above
ambient) up to +250 °C
Voltage / Power Rating
•
230 V (+/- 10%), 50/60 Hz
•
ca. 1.400 W (during heating)
•
adaption of effective heating power
depending on the set temperature
Product specification
Universal oven
Model UFE 400
Standard equipment
Ventilation and Control
•
forced air circulation by quiet air turbine,
adjustable in 10% steps via process controller
•
continuous adjustment of pre-heated fresh air
admixture
•
vent connection with restrictor flap
•
fuzzy-supported PID microprocessor
controller with integrated autodiagnostic
system with fault indication
•
solid state switching unit
•
2 Pt100 sensors Class A in 4-wire-circuit,
mutually monitoring and taking over the
performance at the same temperature value
•
digital 7-day-programme-timer with real time
clock, precise minute setting, for one set
value or ramp operation in combination with:
•
integrated digital timer for tempering profile
with max. 4 segments, each segment
adjustable from 1 min to 999 hours:
DELAYED ON
HEATING UP
HOLD or HOLD set-temperature-dependent
with guaranteed holding time
COOLING DOWN
•
"Loop" function to repeat profile 1-99 times or
endless
•
digital display (LED) of all set parameters,
such as temperature, weekdays, time, fan
speed, ramp segments and set-up values
•
resolution of display for set value 0,5 °C; for
actual value 0,1 °C below 99,9 °C, 0,5 °C
above 100 °C
•
long-term logging (ring store) of all relevant
data, GLP-conforming as data logger - 1024
kB
•
serial interface RS 232 including MEMMERT
Software "Celsius" for programming and
documentation
•
incl. works calibration certificate for +160 °C