CLIMATE CHAMBER Memiliki 3 tipe ukuran volume : 108 L , 256 L, 749 L ; Terdiri dari 3 model: climate chamber ICH 110/260/750 temperature/humidity , ICH 110L/260L/750L dengan cahaya lampu , ICH 110C(260C/750C dengan kontrol CO2 ; Range Temperatur : * - 10 °C s.d. + 60 °C (model ICH) * 0 °C s.d.+ 60 °C (model ICH L and ICH C) ; 4.Twin display ; 5.Kontrol Digital untuk kelembapan 10-80% rH ; 6.Double door (pintu Kaca)
| Temperature | 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
|---|---|
| Humidity | humidity supply with distilled water from external tank by self-priming pump |
| Humidity | humidity supply with distilled water from external tank by self-priming pump |
| Humidity | humidification by hot steam generator |
| Humidity | active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 % |
| ControlCOCKPIT | adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays |
| Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
| from -10°C up to +60°C without humidity | |
| from +10°C up to +60°C with humidity | |
| resolution of display for setpoint and actual temperature values | 0.1°C |
|---|---|
| Function SetpointWAIT | the process time does not start until the set temperature is reached |
|---|---|
| Language setting | German/English/Spanish/French |
| Calibration | three freely selectable values each, temperature/humidity/CO2 |
| adjustable parameters | temperature (Celsius or Fahrenheit), fan speed, relative humidity, programme time, time zones, summertime/wintertime |
| Fan | forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps |
|---|
| Documentation | programme stored in case of power failure |
|---|---|
| Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
| Temperature control | mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature |
|---|---|
| Temperature control | over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display |
| AutoSAFETY | additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature |
| Autodiagnostic system | integral fault diagnostics for temperature and humidity control |
| Alarm | visual and acoustic |
| no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior | |
|---|---|
| energy-saving use of CFC-free cooling/heating system (refrigerant R134a) | |
| highly efficient automatic defrosting system | |
| air jacket heating system for gentle all-around heating |
| Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
|---|---|
| Door | inner glass door |
| Housing | rear zinc-plated steel |
| Internals | 2 stainless steel grids |
| Installation | on lockable castors |
| Max. loading of chamber: 150 kg | |
| w(A) x h(B) x d(C): 560 x 480 x 400 mm | |
| Volume 108 l |
| w(D) x h(E) x d(F): 745 x 1233 x 634 mm |
| Voltage Electrical load | 230 V, 50/60 Hz approx. 1350 W |
|---|---|
| Voltage Electrical load | 115 V, 50/60 Hz approx. 1350 W |
| the appliances must be transported upright | |
|---|---|
| Customs tariff number | 8419 8998 |
| Country of origin | Federal Republic of Germany |
| WEEE-Reg.-No. | DE 66812464 |
| Dimensions approx incl. carton B x H x T: 880 x 1410 x 810 mm | |
| Net weight approx. 109 kg | |
| Gross weight carton approx. 137 kg | |