HIRAYAMA, HV50 HICLAVE WITH 3PCS SS WIRE BASKETS

Prod #HV50

Portable, Top-Loading Autolaves with Advanced Safety and User-Friendly Features

The Most Advanced Autoclaves with Unique Features :

*Electromechanical lock system
*Dual-sensing interlock mechanism
*Double-sensing system for air evacuation
*Programmable auto start
*Process status display
*Process mode for agar

Product Overview

Hirayama HV-50,  50 Liter Upright Autoclave

Desain hemat-ruang dengan tutup terbuka ke atas.
Mulai otomatis yang dapat diprogram untuk memulai siklus hingga 1 minggu kemudian.
Interlock tutup pengindraan ganda mencegah pembukaan tutup saat ruang bertekanan atau ketika suhu melebihi 97 ° C.
Pembersihan udara otomatis oleh mikroprosesor mencegah tekanan berlebih selama siklus.
Tuas sekali sentuh untuk membuka dan menutup tutup ruang dengan mudah.
Laju pembuangan uap yang dapat diprogram.
Sensor penutupan pintu memberikan prompt dan siklus tidak akan dimulai kecuali pintu ditutup dengan benar.
Tampilan dalam proses menunjukkan status dan perkembangan siklus sterilisasi.
Mode sterilisasi agar untuk mensterilkan agar dan kemudian secara otomatis mempertahankannya pada suhu cair.
Aksesori unik: Sensor mengambang untuk memantau suhu beban.
Timer sterilisasi diaktifkan hanya ketika suhu beban mencapai suhu yang disetel.
Mode peleburan agar-agar dengan cepat agar-agar dan mempertahankannya pada suhu cair, atau untuk start-up yang panas dan lebih cepat
Sistem cadangan memori mempertahankan parameter yang diprogram pengguna bahkan jika listrik padam.

 

Specifications

 
HV-50
 
TEMPERATURE  
Operating Range
105ºC to 135ºC
 
Display Range
137ºC
 
Warming Range
45ºC to 60ºC
Agar Dissolution
Heating at 60ºC to 100ºC for up to 1 hour (temperature and time programmable), followed by 45ºC to 60ºC 
(programmable for up to 20 hours to prevent coagulation)
GAUGE PRESSURE  
Operating Range
3 to 30 psi (0.02 to 0.206MPa)
 
Maximum Allowable
38 psi (0.26MPa)
 
TIMER  
Sterilization
1 to 250 minutes
Delayed Auto-start
Program autoclave to start 1 minute to 7 days later
DIMENSIONS  
Overall (W x D x H)
20" x 21" x 41"
(51 cm x 53 cm x 104 cm)
 
Chamber
(Diameter x Height)
50 liters
11.8" x 28.0" (30 cm x 71 cm)
 
SAFETY DEVICES
Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, over-current detector and circuit breaker
PROGRAMS  
1. For agar
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Warming -> Completion
2. For liquids
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Completion
3. For solids
Heating -> Sterilization -> Exhaust (fast, default) -> Completion
4. For melting agar or for faster start-up
Rapid melting at 60°C to 100°C (temperature and time programmable) -> Maintain at 45°C to 60°C (temperature programmable) -> Completion
ACCESSORIES  
Stainless steel
Mesh wire baskets, buckets (pails), and round test tube racks
Factory or field installed
Floating (load) sensor, digital printer, and data logger
POWER  
Source
120V, 50/60 Hz (220V/240V available)
Consumption
2000 watts (20000VA)
 
SHIPPING WEIGHT
180 lb (82 kg)

 

Retrofittable accessories (Any combination possible) :
A : Forced cooling unit
F : Floating sensor
P : Digital printer
W : Automatic water supply
Z : All combined(A+F+P+W)

 