Spesifikasi :
AC Power Adapter and USB/DC Adapter : Included
Automatic Buffer Recognition : Yes
Barometric Pressure Measurement : Automatic compensation of DO when using an LDO or LBOD probe
Calibration : Demal (1D/ 0.1D/ 0.01D);
Molar (0.1M/ 0.01M/0.001M);
NaCl (0.05%; 25µS/cm; 1000µS/cm; 18mS/cm);
Standard sea water;
User defined
Calibration Intervals/Alerts/Reminder : Off, Selectable from 2 hours to 7 days
Compliance Certifications : CE.WEEE
Conductivity measurement : Yes
Conductivity measurement : Temperature correction:None; Linear; NaCl Non-Linear Natural Water.
Conductivity Measurement Range : 0.01 µS/cm - 200.0 mS/cm
Conductivity resolution : 0.01 µS/cm - 0.1 mS/cm upon selected measuring range
Contents : No
Custom Calibration Standards : Yes
Data Export : Download via USB connection to PC or flash memory device. Automatically transfer entire data log or as readings are taken.
Data Memory : 500 records/FIFO
Data Storage : Automatic, GLP ISO compliant reading data stored with calibration details.
Digital (intelligent) electrode inputs : 2 channels
Display : Detailed mode/Large mode
Display Lock Function : Continuous / Auto-stabilization ("press to read") / At Interval
Display Type : 240 x 160 pixel LCD with backlight illumination
DO Measurement Range : 0.1 - 20.0 mg/L (ppm) 1 - 200% saturation
DO Resolution : 0.1
DO sensor calibration : * 100% (water-saturated air (100%) calibration
* 100% with 0 (water-saturated air (100%) calibration with 0 point
* mg/L (calibration with a specified dissolved oxygen concentration (mg/L) solution)
* mg/L with 0 (calibration with a specified dissolved oxygen concentration (mg/L) solution with 0 point)
* Factory (calibration with the default LDO calibration)
GLP Features : Date; Time; Sample ID; Operator ID
Inputs : 2
Instrument : Portable
IP Rating : IP67
ISE Direct Measurement Range : Yes
ISE Electrode Calibration : 2 - 5 points
Languages : English, French, German, Italian, Spanish, Danish, Dutch, Polish, Portuguese, Turkish, Swedish, Czech, Russian
Measurement Method : Probe specific programmed method settings
Model : HQ40D – Multi/2 Channels
mV Measurement at Stable Reading : Yes
mV Measurement Range : -1500 - 1500 mV
mV Resolution : 0.1 mV
Needed Cable Length : 1
Operating Error Messages : Clear text error messages displayed
Operating Humidity : 90 % relative humidity (non-condensing)
Operating Interface : Soft Touch Keypad
Operating Temperature : 0 - 60 °C (32 - 140 °F)
ORP Electrode Calibration : Predefined ORP standards (including Zobell's solution)
Parameter : pH/Oxydo Reduction Potential (ORP)
Conductivity/Total Dissolved Solid (TDS)/Salinity/Resistivity
Dissolved Oxygen (DO)
Biochemical Oxygen Demand (BOD)
Ion Selective Electrode (ISE) : Ammonia, Ammonium, Chloride, Fluoride, Nitrate, Sodium
Parameters : pH
ORP
NA
PC Data Transfer Software : HQD Series Meter Data Transfer Utility
pH Buffer Sets : Color-coded: 4.01, 7.00, 10.01 pH;
IUPAC : 1.679, 4.005, 7.000, 10.012, 12.45
DIN : 1.09, 4.65, 9.23
User-defined custom buffer sets
pH Electrode calibration : 1 - 3 Calibration points
Calibration summary data logged and displayed
pH Measurement Range : 0 - 14 pH
pH Resolution:Selectable:
0.001/0.01/0.1 pH
Printer : Yes , Optional
Probes included : PHC20101, MTC10101
Probe Type : Intellical Standard Laboratory or Rugged Field
Region : US
Salinity Measurement Range : 0 - 42 (ppt) (‰)
Salinity Resolution : 0.01 (ppt) (‰)
Sensor A : PHC20101
Sensor B : MTC10101
Sensor C : NA
Sensors:Sensor A : PHC20101, MTC10101
Sensor B : MTC10101
Sensor C:
Simultaneous measurements : Yes, 2 channels
TDS Measurement Range : 0.00 mg/L - 50.0 g/L NaCl
TDS Resolution : 0.01 mg/L up - 0.1 g/L upon measuring range
Temperature Compensation : Automatic Temperature compensation for pH
Automatic Temperature compensation for pH
Temperature Measurement : °C or °F
Temperature Resolution : 0.1
Type of Measurement : QC Lab
Weight : 335 g (0.75 lb) without batteries; 430 g (0.95 lb) with
includes : HQ40D Portable Meter; 4 AA batteries; Power adapter, USB/DC power adapter for data transfer; quick-start guide and user manual; HQD Meter Stand; Probe Stand; Buffer Solution, pH 4.01/7.00/10.01 (NIST), color-coded red, 500 mL; ZoBell's ORP/Redox Standard Solution, 500 mL; pH Electrode Storage Solution, 500 mL.
Intellical PHC201 Laboratory pH electrode with storage soaker bottle, 1 m cable, calibration certificate, and Basic User Manual.
Intellical MTC101 Laboratory ORP/RedOx electrode with storage soaker bottle, 1 m cable, test certificate, and Basic User Manual.