Specifications :
Accuracy : ±25% of set point value
Alarm : 2 consectutive cycles above set point activate alarm. 1 cycle below set point cancels alarm.
Alarm points : 0.3 mg/L of CaCO3 trip point
Certifications : NRTL certified to UL and CSA standards, and CE approved
Cycle Time : 2 average
Dimensions (H x W x D) : 419 mm x 318 mm x 178 mm
Fitting Type :
Inlet: ¼-in. OD tube quick-disconnect fitting
Outlet: ½-in. ID flexible hose
Light Source : LED with peak wavelenght of 610 nm
Maintenance Interval :
Standardise and replenish reagents every 2 months using EDTA or high-level sample
Replace pump tubes every 6 months
Material Enclosures : ABS plastic, large plastic windows to view alarm and reagents level
Measurement Method : Colorimetric
Operating Humidity : 5 - 95%, non-condensing
Operating Temperature Range : 5 - 40 °C
Output : 1 SPDT relay
Power Requirements (Hz) : 50/60 Hz
Power Requirements (Voltage) : 115/230 V AC
Reagent Consumption : 500 ml in each indicator and buffer every 2 months
Reagent Requirements : each of indicator and buffer. Replenish every two months for continuous operation.
Repeatability :
± 10 % of set point value in 0.3 - 2 mg/L ranges
± 4 % of set point value in 5 - 100 mg/L ranges
Sample Flow Rate : 50 - 500 mL/min
Sample Inlet Connection : 1/4-in. OD tube quick-disconnect fitting
Sample Pressure :
0.07 - 0.34 bar (0.10 bar is optimum)
Sample conditioning : 0.10 - 5.17 bar
Service Interval :
Standardize and replenish reagents every 2 months using EDTA or high-level sample
Replace pump tubes every 6 months
Storage Conditions : -40 - 60 °C
Weight : 11.3 kg