CLF10sc Free Chlorine Analyzer mengukur klorin bebas secara continous selama proses kontrol berjalan. Hal ini didasarkan pada teknologi amperometri dan dilengkapi dengan controller digital SC200. CLF10sc dapat digunakan dalam sebagian besar aplikasi industri. Analyzer ini dilengkapi dengan sensor pH. -Kompatibel dengan \"Plug and Play\" Hach Digital Controller. -Tidak ada Penggantian Reagen, tidak ada limbah . -Sesuai EPA . -Diagnostik yang canggih. -Proses Monitoring dan Pengendalian
Spesifikasi :
|
Accuracy |
Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit) |
|
Accuracy 2 |
Free Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration) |
|
Accuracy 3 |
Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration) |
|
Accuracy 4 |
Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5 |
|
Cable Connection |
Cable Connection |
|
Cable Length |
1 m (between gateway and sc controller) |
|
Calibration Method |
1-point or 2-point (zero and slope) calibration |
|
Certifications |
CE compliant for conducted and radiated emissions: |
|
Connection Drain Line |
pH Flow Cell Outlet: 1/2-inch ID |
|
Connections |
1/4 inch OD |
|
Controller Compatibility |
sc controller platform |
|
Diameter |
25 mm |
|
Dimensions |
mm x 25 mm (7.68 in x 0.98 in) |
|
Fitting type |
USA |
|
Interferences |
Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits. |
|
Length |
Sensor: 195 mm |
|
Lower Limit of Detection (LOD) |
30 ppb (0.03 ppm) |
|
Material |
Corrosion-resistant materials, fully-submersible |
|
Measurement Method |
Reagentless, electrochemical, three-electrode amperometric system |
|
Measurement Range 2 |
0 to 20 ppm |
|
Mounting |
Panel or Wall Mounting |
|
Operating Temperature Range |
0 to 45 °C |
|
Parameter |
Free Chlorine |
|
pH Range |
4 - 9 |
|
Power Requirements (Voltage) |
110 - 230 V AC |
|
Pressure Range |
0.5 bar, no pressure impulses and/or vibrations |
|
Repeatability |
30 ppb or 3%, whichever is greater |
|
Resolution |
0.001 ppm (1 ppb) |
|
Response Time |
Free Chlorine: 140 s or less for 90% change (T90) |
|
Sample Flow Rate |
30 to 50 L/h |
|
Sample Temperature |
5 to 45 °C (41 to 113 °F) |
|
Storage Conditions |
-20 °C to 60 °C dry |
|
Warranty |
1 Year |
|
Weight |