Specifications :
Accuracy :
Free Chlorine: ±3% of the reference test (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2 :
Free Chlorine: ±10% of the reference test (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3 :
Total Chlorine: ±10% of the reference test (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4 :
Total Chlorine: ±20% of the reference test (DPD) at stable pH greater than 8.5
Cable Connection : 5 pin, M12 connector
Cable Length : 1 m (between gateway and sc controller)
Calibration Method : 1-point or 2-point (zero and slope) calibration
Certifications :
CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Additional certifications when connected to an sc controller
Connection Drain Line : 1/2 in ID
Connections : 1/4 in OD
Controller : Panel only
Controller Compatibility : sc controller platform
Description : CLF10sc Free Chlorine Analyzer (Panel Only), Grab Sample
Dimensions :
mm x 25 mm (7.68 in x 0.98 in)
Sensor Only: 195
Fitting type : USA
Includes : CL10sc Panel, 1 m digital extension cable, Panel Manual, Chlorine Sensor Manual.
Interferences :
Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits.
Total chlorine: chlorine dioxide, ozone and chalk deposits
Length : Sensor: 195 mm
Lower Limit of Detection (LOD) : 30 ppb (0.03 ppm)
Material :
Corrosion-resistant materials, fully-submersible
(stainless steel, PEEK, PVC, silicon rubber and polycarbonate)
Measurement method : Reagentless, electrochemical, three-electrode amperometric system
Mounting : Panel or Wall Mounting
Operating temperature range : 0 - 45 °C
Options : Panel Only
Parameter : Free Chlorine
pH Monitoring Required? : None
pH Range : 4 - 9
Power requirements (Voltage) : 24 V DC
Pressure Range : 0.5 bar, no pressure impulses and/or vibrations
Range : 0 - 20 ppm Cl?
Repeatability : 30 ppb or 3%, whichever is greater
Response time :
Free Chlorine: 140 s or less for 90% change (T90) at a stable temperature and pH
Response Time :
Total Chlorine: 100 s or less for 90% change (T90) at a stable temperature and pH
Sample Flow Rate :
30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature : 5 - 45 °C (41 - 113 °F)
Storage conditions : -20 °C to 60 °C dry
Weight :
Approximately 5.5 kg (12 lbs)
Panel and empty panel-mounted components only
What's included? :
CL10 sc Panel, 1 m digital extension cable, Panel Manual, Chlorine Sensor Manual.